技術簡介
OCT系統掃描時,為方便看到整個眼底視網膜之影像而開發,本系統利用低強度紅外光作為聚焦光源,掃描過程中不會使瞳孔縮小,無需散瞳劑。此外,本系統全眼底掃描過程花費時間比起眼球轉動時間更短,因此不會造成受測者的不適。
Abstract
SLO is designed to search for the image of the entire retina when the OCT system is scanned. This SLO system utilizes a low intensity infrared light as the source of focused light beam. Without mydriatic, the scanning process does not cause miosis. The whole retinal scanning process takes time shorter than eye movements, it does not cause any discomfort.
技術規格
波長:780 nm (Nominal)
解析度:13m
焦深:0.25mm
掃描角度:30° x 30°/ 20° x 20°/ 15° x 15°
數位影像大小:768 x 768 / 512 x 512 / 256 x 256
橫向解析度:11um/pixel digital (at normal incidence)
影像擷取頻率:6Hz / 9Hz / 12 Hz
調焦範圍: -12 diopters to + 12 diopters spherical
Technical Specification
Wavelength 780 nm (Nominal)
Optical resolution 13m
Depth of focus 0.25mm
Scan angle 30° x 30°/ 20° x 20°/ 15° x 15°
Digital image size(pixel xpixel) 768 x 768 / 512 x 512 / 256 x 256
Lateral resolution:11um/pixel digital (at normal incidence)
Image acquisition frequency:6Hz / 9Hz / 12 Hz
Focus adjustment range:-12 diopters to + 12 diopters spherical
技術特色
快速掃描、免散瞳劑、精密光機引擎設計組裝能力,可依需求客製化製作
高速影像擷取(12Hz)可避免眼球微顫動影響影像品質
應用範圍
視網膜眼底觀察、黃斑部病變、視神經變化
接受技術者具備基礎建議(設備)
自動準直儀、二維光形量測儀、CNC工具機等
接受技術者具備基礎建議(專業)
光學設計、光路調校、光/機設計、精密加工、機電整合、人機儀控
聯絡資訊
聯絡人:張奇偉 醫電與診斷技術中心
電話:+886-3-5917450 或 Email:ChirWeeiChang@itri.org.tw
客服專線:+886-800-45-8899
傳真:+886-3-5829781